When to use automatically estimated or manually measured IRF on MicroTime 200?

When fitting TCSPC data measured with a MicroTime 200 (MT200), SymPhoTime 64 (SPT) offers two options for defining the Instrument Response Function (IRF):

  • Automatically estimated IRF: SymPhoTime estimates the IRF directly from the rising edge of the decay curve.

  • Manually measured IRF: A separately recorded scatter measurement is used as the IRF.


Automatically estimated IRF

When to use it

  • The sample lifetime is significantly longer than the timing resolution of the system — roughly >700 ps as a guideline.

  • You are performing a quick, exploratory analysis and do not need maximum accuracy.

  • No scatter measurement was recorded during the experiment.

Limitations

  • Not suitable for very short lifetimes (below ~700 ps) where the IRF significantly overlaps with the decay.

  • Less reliable for multi-exponential decays with a fast component.

  • Not suitable for anisotropy measurements.


Manually measured IRF

When to use it

  • The sample has a very short lifetime (<700 ps), where the IRF has a significant contribution.

  • You need accurate results for multi-exponential decays with a fast component.

  • You are performing anisotropy measurements.

  • Highest accuracy is required.

Requirements

  • A scatter measurement must have been recorded during the experiment (e.g., using a scattering solution such as Ludox).

  • The scatter measurement must use the same optical path (objective, filters, pinhole settings).


Practical recommendation

Use the manually measured IRF whenever maximum accuracy is needed: short lifetimes (<700 ps), multi-exponential decays with fast components, or anisotropy measurements. In all other cases, the automatically estimated IRF is a convenient and reliable choice.